How to cite this article: Borges LP, De Oliveira MGB, de Souza DRV, Dos Santos KA, Heimfarth L, et al. Can the Wondfo® SARS-CoV-2 IgM/IgG antibodies be used as a rapid diagnostic test? Arch Biotechnol Biomed. 2020; 4: 013-017.

DOI: 10.29328/journal.abb.1001017

Copyright License: © 2020 Borges LP, et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Keywords: COVID-19; New coronavirus; Immunochromatographic assay; Validation; Fluorescence immunoassay test

 FullText PDF

Can the Wondfo® SARS-CoV-2 IgM/IgG antibodies be used as a rapid diagnostic test?

Lysandro Pinto Borges1, Makson Gleydson Brito De Oliveira2*, Daniela Raguer Valadão de Souza3, Kezia Alves Dos Santos1 and Luana Heimfarth4

1Department of Pharmacy, Federal University of Sergipe, Marechal Rondon Avenue, S/N, Rosa Elza, CEP: 49.000-100, São Cristóvão, SE, Brazil
2Department of Medicine, Federal University of Sergipe, Lagarto, SE, Brazil
3Department of Education and Health, Federal University of Sergipe, Lagarto, SE, Brazil
4Department of Physiology, Federal University of Sergipe, Marechal Rondon Avenue, S/N, Rosa Elza, CEP: 49.000-100, São Cristóvão, SE, Brazil

*Address for Correspondence: Makson Gleydson Brito de Oliveira, PhD, Professor, Medicine Department, Federal University of Sergipe, Av. Governador Marcelo Déda, 13, CEP 49400-000 Lagarto, Sergipe, Brazil, Tel: 55 (79) 36322080. Email: makson_gbo@hotmail.com

Background: An outbreak of novel coronavirus (SARS-CoV-2) disease (COVID-19) has rapidly spread worldwide. The aim of this study was to evaluate and validate the performance of the Wondfo® lateral-flow immunochromatographic assay that detect SARS-CoV-2- IgG, IgM antibodies (Wondfo® IC), using the results obtained by the fluorescence immunoassay test as reference diagnostic.

Material and methods: 97 serum specimens collected and analyzed by four independent laboratories of Sergipe/Brazil was used for validated the Wondfo® SARS-CoV-2 IgM/IgG antibodies test. The COVID-19 positive serum specimens were determined by fluorescence immunoassay technique, used as reference standard.

Results: An overall of 97 serum specimens show 39 (39/97) SARS-CoV-2 IgG positive specimens, 33 (33/97) SARS-CoV-2 IgM positive specimen and 25 non-reagent specimens (25/97). However, the Wondfo® IC assay detected only 9 (9/97) IgM/IgG positive specimen and 25 (25/97) no-reagent specimen. A weak correlation was found between the outcomes of the Wondfo® IC assay and fluorescence test. The accuracy between the two tests was 32.08%. The sensitivity, specificity, positive predictive value, and negative predictive value of Wondfo® IC assay were of 11.12%, 100%, 100% and 25.27%, respectively. Moreover, no false positive sample was determinate, whereas 88.89% of false negative results were found.

Conclusion: The Wondfo® IC test failed in providing a quick, valid, and reliable results and appears not to be a good alternative for clinical use in detecting pandemic coronavirus. However, if the limitations of the rapid test are known, some correction factors can be used in order to adjust the epidemiological data.

An outbreak of novel coronavirus (SARS-CoV-2) disease (COVID-19) was first identified in Wuhan City, Hubei Province, China in December 2019 and has rapidly spread worldwide, since been declared a pandemic by the World Health Organization (WHO) [1]. In the May 2020, COVID-19 has inflicted more than 4 million people globally with about 307.537 death [2]. Most people that are infected by SARS-CoV-2 present mild or no symptoms, but some COVID-19 patients develop severe pneumonia, acute respiratory distress syndrome (ARDS), multiorgan failure that could evolve to death [3].

Adequate diagnosis of SARS-CoV-2 infection is crucial for define the therapeutic management of patients, the establishment of infection control protocols and prevent the dissemination of the virus to new communities. Quantitative reverse-transcription polymerase chain reaction (RT-PCR) analysis for SARS-CoV-2 RNA is considered the gold standard for detecting COVID-19 disease, however the sensitivity of this method may vary by the time of infection and the viral load [4,5], as well as the RT-PCR is an expensive assay because require special equipment and reagents. Therefore, an alternative laboratorial test was necessary. In this context, the detection of IgG, IgM and IgA antibodies against the SARS-CoV-2 have been gain importance in the diagnosis of COVID-19 due to its rapid and reliable diagnostic test. The serological test become essential for epidemiological information and contribute to adequate lockdown exit strategies and vaccine development [6,7].

One serological test that have been obtained notoriety in the diagnostic of COVID-19 is the immunochromatographic (IC) assay for IgM and IgG antibodies against the virus. The IC assay is a rapid and high-throughput method for diagnosing viral infections and it is accepted as a point-of-care test [8]. Recently, various commercial IC assays were developed for COVID-19 diagnostic, and these tests have been used in clinical setting. However, their efficacy and clinical usefulness need to be evaluated and validate. An appropriate diagnostic method for a disease must have a high sensibility and specificity, as well as a good performance under various conditions [9].

Therefore, the aim of this study was to evaluate and validate the performance of the Wondfo® lateral-flow immunochromatographic assay that detect SARS-CoV-2 IgG, IgM antibodies, comparing with the results obtained by the fluorescence immunoassay test (FIA). The clinical detection sensitivity and specificity of Wondfo® IC test were measured using blood samples collected from 72 FIA confirmed COVID-19 patients and 25 negative patients at 4 different laboratories.

Study design

This study included 97 residual human sera obtained by venous blood from patients and healthy people of four laboratories of Sergipe/Brazil, Lab1(UFS): 31 samples; Lab2 (CIRURGIA Laboratory): 25 samples; Lab3 (CLIMED Laboratory): 21 samples; Lab4 (SOLIM Laboratory): 20 samples. The SARS-CoV-2 IgM or IgG specimen were diagnosed positive using a fluorescence immunoassay technique (FIA). We use a FIA with a semi-quantitative detection of IgG and IgM antibodies for SARS-COV-2 in the semi-automatic device iChroma2 by Bodytech (south korea) Samples were divided in COVID-19 reagent and non-COVID-19 reagent based in FIA COI value, COI<0.9 is detected as non-reagent and COI >1.1 is detected as reagent.

The sample was collected in April 2020 by blood puncture in tubes with separating gel, centrifuged at 3000 RPM for 10 min and the serum was separated to SARS-CoV-2 IgM or IgG antibody analysis. The FIA assay and the immunochromatographic test were performed on the same day of collection following the manufacturing protocol. All Wondfo® SARS-CoV-2 IgM/IgG antibody test was performed and analyzed by a blinded operator. For kit precision/reproducibility study, 30 specimens were blinded and randomized chosen, and the Wondfo® IC assay was performed in triplicate.

The inclusion criterium require non reagent diagnostic for influenza virus, Influenza A, Influenza B, HIV, Epstein-bar virus, hepatitis B and hepatitis C antibody.

Wondfo® Lateral-flow immunochromatographic assay that detect SARS-CoV-2 IgM/IgG antibodies (Wondfo® IC test)

Wondfo® SARS-CoV-2 IgM/IgG antibody test is an immunochromatographic assay for rapid and quantitative detection of SARS-CoV-2 IgM/IgG antibody in human biological samples. Venous blood serum was subjected to the SARS-CoV-2 IgM/IgG antibody test, using lateral flow method assay in accordant with the manufacturer’s protocol (Guangzhou Wondfo Biotech). In brief, 10 μL of serum specimen were added onto the sample loading area followed by 80 μL (2 drops) of buffer. After 15 min of incubation, viral IgM- or IgG-containing positive samples could show up both the T line (test) and C line (control); the samples with only C line were regarded as negative.

Accuracy= NTP+ NTN (NTP+FN+ FP+NTP)  MathType@MTEF@5@5@+=feaaguart1ev2aaatCvAUfeBSjuyZL2yd9gzLbvyNv2CaerbuLwBLnhiov2DGi1BTfMBaeXatLxBI9gBaerbd9wDYLwzYbItLDharqqtubsr4rNCHbGeaGqk0Jf9crFfpeea0xh9v8qiW7rqqrFfpeea0xe9Lq=Jc9vqaqpepm0xbba9pwe9Q8fs0=yqaqpepae9pg0FirpepeKkFr0xfr=xfr=xb9adbaqaaeGaciGaaiaabeqaamaabaabaaGcbaaeaaaaaaaaa8qacaqGbbGaae4yaiaabogacaqG1bGaaeOCaiaabggacaqGJbGaaeyEaiabg2da9maalaaapaqaa8qacaWGobGaamivaiaadcfacqGHRaWkcaGGGcGaamOtaiaadsfacaWGobaapaqaa8qacaGGOaGaamOtaiaadsfacaWGqbGaey4kaSIaamOraiaad6eacqGHRaWkcaGGGcGaamOraiaadcfacqGHRaWkcaWGobGaamivaiaadcfacaGGPaGaaiiOaaaaaaa@5649@

Sensibility = NTP (NTP+FN) MathType@MTEF@5@5@+=feaaguart1ev2aaatCvAUfeBSjuyZL2yd9gzLbvyNv2CaerbuLwBLnhiov2DGi1BTfMBaeXatLxBI9gBaerbd9wDYLwzYbItLDharqqtubsr4rNCHbGeaGqk0Jf9crFfpeea0xh9v8qiW7rqqrFfpeea0xe9Lq=Jc9vqaqpepm0xbba9pwe9Q8fs0=yqaqpepae9pg0FirpepeKkFr0xfr=xfr=xb9adbaqaaeGaciGaaiaabeqaamaabaabaaGcbaaeaaaaaaaaa8qacaqGtbGaaeyzaiaab6gacaqGZbGaaeyAaiaabkgacaqGPbGaaeiBaiaabMgacaqG0bGaaeyEaiaabckacqGH9aqpdaWcaaWdaeaapeGaamOtaiaadsfacaWGqbaapaqaa8qacaGGOaGaamOtaiaadsfacaWGqbGaey4kaSIaamOraiaad6eacaGGPaaaaaaa@4DA7@

Specificity= NTN (FP+ NTN) MathType@MTEF@5@5@+=feaaguart1ev2aaatCvAUfeBSjuyZL2yd9gzLbvyNv2CaerbuLwBLnhiov2DGi1BTfMBaeXatLxBI9gBaerbd9wDYLwzYbItLDharqqtubsr4rNCHbGeaGqk0Jf9crFfpeea0xh9v8qiW7rqqrFfpeea0xe9Lq=Jc9vqaqpepm0xbba9pwe9Q8fs0=yqaqpepae9pg0FirpepeKkFr0xfr=xfr=xb9adbaqaaeGaciGaaiaabeqaamaabaabaaGcbaaeaaaaaaaaa8qacaqGtbGaaeiCaiaabwgacaqGJbGaaeyAaiaabAgacaqGPbGaae4yaiaabMgacaqG0bGaaeyEaiabg2da9maalaaapaqaa8qacaWGobGaamivaiaad6eaa8aabaWdbiaacIcacaWGgbGaamiuaiabgUcaRiaacckacaWGobGaamivaiaad6eacaGGPaaaaaaa@4D93@

Positive predictive value = NTP (NTP+FP) MathType@MTEF@5@5@+=feaaguart1ev2aaatCvAUfeBSjuyZL2yd9gzLbvyNv2CaerbuLwBLnhiov2DGi1BTfMBaeXatLxBI9gBaerbd9wDYLwzYbItLDharqqtubsr4rNCHbGeaGqk0Jf9crFfpeea0xh9v8qiW7rqqrFfpeea0xe9Lq=Jc9vqaqpepm0xbba9pwe9Q8fs0=yqaqpepae9pg0FirpepeKkFr0xfr=xfr=xb9adbaqaaeGaciGaaiaabeqaamaabaabaaGcbaaeaaaaaaaaa8qacaqGqbGaae4BaiaabohacaqGPbGaaeiDaiaabMgacaqG2bGaaeyzaiaabckacaqGWbGaaeOCaiaabwgacaqGKbGaaeyAaiaabogacaqG0bGaaeyAaiaabAhacaqGLbGaaeiOaiaabAhacaqGHbGaaeiBaiaabwhacaqGLbGaaeiOaiabg2da9maalaaapaqaa8qacaWGobGaamivaiaadcfaa8aabaWdbiaacIcacaWGobGaamivaiaadcfacqGHRaWkcaWGgbGaamiuaiaacMcaaaaaaa@5B23@

Negative predictive value= NTN (NTN+FN) MathType@MTEF@5@5@+=feaaguart1ev2aaatCvAUfeBSjuyZL2yd9gzLbvyNv2CaerbuLwBLnhiov2DGi1BTfMBaeXatLxBI9gBaerbd9wDYLwzYbItLDharqqtubsr4rNCHbGeaGqk0Jf9crFfpeea0xh9v8qiW7rqqrFfpeea0xe9Lq=Jc9vqaqpepm0xbba9pwe9Q8fs0=yqaqpepae9pg0FirpepeKkFr0xfr=xfr=xb9adbaqaaeGaciGaaiaabeqaamaabaabaaGcbaaeaaaaaaaaa8qacaqGobGaaeyzaiaabEgacaqGHbGaaeiDaiaabMgacaqG2bGaaeyzaiaabckacaqGWbGaaeOCaiaabwgacaqGKbGaaeyAaiaabogacaqG0bGaaeyAaiaabAhacaqGLbGaaeiOaiaabAhacaqGHbGaaeiBaiaabwhacaqGLbGaeyypa0ZaaSaaa8aabaWdbiaad6eacaWGubGaamOtaaWdaeaapeGaaiikaiaad6eacaWGubGaamOtaiabgUcaRiaadAeacaWGobGaaiykaaaaaaa@59DA@

Where, NTP, number of true positive; NTN, number of true negative; FP, number of false positive; FN, number of false negative

Ethical statement

The study was approved by the Research Ethics Committee of the Federal University of Sergipe, nº CAAE 31018520.0.0000.5546, April 24, 2020.

IgM and IgG antibodies for SARS-CoV-2 could be detected by lateral-flow immunochromatographic assay (IC) and this rapid test have been currently used for the diagnostic of COVID-19 patients. In this study we compare the result obtain by Wondfo® IC assay (Wondfo® SARS-CoV-2 antibody test) and results obtain from fluorescence immunoassay (FIA), taken as reference standard.

An overall of 97 serum samples were collected and analyzed by four different and independent laboratories (Lab1: n=31; Lab2: n=25; Lab3: n=21; Lab4: n=20) through fluorescence immunoassay technique (FIA) and the results show 39 (39/97) SARS-CoV-2 IgG positive specimen (COI 2.2-53.2), 33 (33/97) SARS-CoV-2 IgM positive specimen (COI 1.3-5.0) and 25 non-reagent specimen (COI <0.9) (25/97) (Table A), accounting an 74.22% (72/97) SARS-CoV-2 IgG/IgM positive specimens detected by the reference standard technique. The FIA findings were used by define SARS-CoV-2 IgG or IgM positive or negative specimen.

Table A: IC assay and FIA assay findings for patients with COVID-19.
  FIA (n = 97) Wondfo® SARS-CoV-2 antibody test (n = 97)
  Number
Positive /Total
% Number
Positive /Total
%
IgM 33 / 97 34.0    
IgG 39 / 97 41.2    
Total IgG/IgM 72 / 97 74.2 9 / 97 9.3
Non-reagent 25 / 97 27.8 25/97 25.8
FIA: fluorescence immunoassay

Afterward, the same 97 samples were analyzed by Wondfo® COVID-19 IgM/IgG rapid immunochromatographic test (Wondfo® IC) and the results were summarized in table A. Wondfo® IC assay captures total SARS-CoV-2 immunoglobulin (IgG and IgM). All total sample size analyzed at Wondfo® IC assay was successfully performed, because the line in the control zone was clearly visible after 15 min of migration after the specimen was inserted in the cassette.

The Wondfo® IC results show that SARS-CoV-2 IgG/IgM antibody was detected in 9 (9.3%) of the 97 serum specimens collected. As expected, all 25 non-COVID-19 people were negative in Wondfo® IC test. A weak correlation was found between the outcomes of the Wondfo® IC assay (9.3% of SARS-CoV-2 IgM/IgG positive samples) and fluorescence test (74.22% of SARS-CoV-2 IgM/IgG positive samples) (Table A). Table B and C represent a descriptive result of each specimen with the individual data of Wondfo® IC test result, comparing COI number obtained in FIA.

Table B: IC assay findings for patients with COVID-19 determined by FIA assay with SARS-CoV-2 IgG positive diagnostic (n=39).
 Lab1  Lab2  Lab3  Lab4
FIA
(COI)
Wondfo®
IC test
FIA
(COI)
Wondfo®
IC test
FIA
(COI)
Wondfo®
IC test
FIA
(COI)
Wondfo®
IC test
29.5 NR 37.3 NR 9.4 NR 26.3 R
29.6 NR 6.2 NR 12.3 NR 10.5 NR
32.0 R 2.2 NR 2.7 NR 8.1 NR
34.0 NR 14.6 NR 11.2 NR 36.8 R
36.0 NR 13.4 NR 21.1 NR 26.7 R
36.4 NR 24.5 NR 5.1 NR 27.8 R
36.9 NR 46.9 NR 36.6 NR 12.2 R
37.2 NR 37.7 NR 9.0 NR 7.7 NR
47.0 NR 50.3 NR 3.4 NR 6.8 R
49.0 NR 53.2 NR     5.4 NR
FIA: fluorescence immunoassay; R: IgG reagent; NR: non-reagent; COI: Cut off index.
FIA cut off index: COI<0,9 non-reagent, COI >1.1 reagent
Table C: IC assay findings for patients with COVID-19 determined by FIA assay with SARS-CoV-2 IgM positive diagnostic (n=33).
 Lab1  Lab2  Lab3  Lab4
FIA
(COI)
Wondfo®
IC test
FIA
(COI)
Wondfo®
IC test
FIA
(COI)
Wondfo®
IC test
FIA
(COI)
Wondfo®
IC test
2.0 R (+) 1.8 NR 1.6 NR 2.4 NR
2.4 NR 1.6 NR 2.4 NR 2.1 NR
2.6 NR 5.0 NR 4.1 NR 2.9 NR
2.8 NR 2.8 NR 2.9 NR 2.9 NR
3.1 NR 2.2 NR 1.9 NR 2.7 NR
3.2 NR 2.2 NR 1.3 NR    
3.4 R(+++) 2.6 NR 1.8 NR    
4.0 NR 1.3 NR 1.6 NR    
4.0 NR 2.0 NR        
4.3 NR 2.8 NR        
FIA: fluorescence immunoassay; R: IgM reagent; NR: non-reagent; COI: Cut off index.
+ represent a weak reaction; +++ represent a strong reaction.
FIA cut off index: COI<0,9 non-reagent, COI >1.1 reagent

Table D represent the performance characteristic of the Wondfo® IC assay compared to the fluorescence assay, taken as reference standard, on the 97 serum specimens. The accuracy between the two tests was 32.08 %. The sensitivity, specificity, positive predictive value and negative predictive value of Wondfo® IC assay was of 11.12%, 100%, 100% and 25.27%, respectively. Moreover, no false positive sample was determinate, whereas 88.89% of false negative results were found (Table D). The analysis shows that Wondfo® IC assay is reproductively (data not shown).

Table D: Accuracy of rapid immunochromatographic tests in 97 specimens of SARS-CoV-2 IgG/IgM reagent and non-reagent cases classified according to the reference test.
  Wondfo® SARS-CoV-2 antibody test
Sensibility 11.12%
Specificity 100%
Positive predictive value 100%
Negative predictive value 25.27%
False-positive 0.0%
False-negative 88.89%
Accuracy 32.08%

In order to confirm the previous findings, an additional test was performed using specimen that have the rRT-PCR for RNA viral positive. The RT-PCR is considered the gold standard assay for COVID-19 diagnostic. The fluorescence assay results show that the specimen was IgM reagent (SARS-CoV-2 IgM positive) (COI 1.4), whereas in Wondfo® IC test, the same sample was detected as non-reagent.

The tracking of SARS-CoV-2 virus in the population help the control of the epidemic situation and facilitate the diagnostic of new cases, becoming an important tool for public health [4]. The adequate diagnosis test is essential for obtain effective and reliable results. This ensures laboratory findings can be traced and patients identified for orientation, isolation, and treatment. Currently recommendation of the WHO for COVID-19 diagnosis indicate the use of molecular tests targeting SARS-CoV-2 virus RNA. However, due to RT-PCR infrastructure limitations and lack of supplies, which limit the number of people with access of a diagnostic tests, a rapid serologic assay was develop to expand laboratories testing capacity and reach all the population [10]. In this line, several rapid tests based on immunochromatographic method for SARS-CoV-2 IgM/IgG antibody have been developed [8,11]. Nevertheless, to achieve the goal of help in public health, this test needs to be a well-validated diagnostic tool that are sensitive, rapid and specific for the detection of SARS-CoV-2.

Demey, et al. [11] using 4 different immunochromatographic rapid tests describe that this kind of diagnostic tool have good performance for the detection of antibodies after SARS-CoV-2 infection. However, Vásárhelyi, et al. [12] found low efficacy in a rapid immunochromatographic tests detecting IgM and IgG antibodies against SARS-CoV-2 virus, suggesting that this test should not be used in the differential diagnosis of coronavirus infection. So, the controversial literature about immunological rapid test lead to the necessity of validation of the SARS-CoV-2 IgM/IgG diagnostic kits in the target population for the purpose of improve the quality of the analysis.

In the present study we evaluated the Wondfo® IC test (Wondfo® SARS-CoV-2 antibody test), trying to validate this immunochromatographic assay. The results obtained by fluorescence immunoassay test (FIA), performed by 4 independents laboratory, was used as reference standard. The analytical results of a commercial Wondfo® IC test kit and the findings obtained by fluorescence test for patients with COVID-19 was compared. The Wondfo® IC assay showed low sensitivity (11.12%) and high specificity (100%) for COVID-19, with a huge number of false-negatives results. The fluorescence assay detected 72 IgG/IgM antibodies specimens, whereas the Wondfo® IC assay just reproduce the same result in 9 of this samples, obtaining a low accuracy. Moreover, the Wondfo® IC assay also failed in detected the SARS-CoV-2 IgM protein in additional specimen confirmed positive for SARS-CoV-2 virus with real-time RT-PCR, set as the gold standard. All this results together, suggest that the Wondfo® IC test did not present a good agreement with the reference standard test or with the RT-PCR assay.

Furthermore, the low negative predictive values indicate that Wondfo® IC kit stumble in detecting the presence of SARS-CoV-2 IgM and IgG against virus infection in several specimen, becoming not suitable for screening COVID-19 infection in the general population. So, based in low sensitivity and elevated negative predictive values, we could propose that the Wondfo® IC test showed low “validity¨.

It is important to note that in this study we used specimen collected by blood puncture in tubes with separating gel, centrifuged quickly and SARS-CoV-2 IgM or IgG serological tests were done on the same day. It is worth remembering that samples cooled and kept for several days can impair the detection of antibodies. Therefore, the prolonged time to perform the analysis after the collection of the specimen could affect the result of the immunological test, due to the degradation of the reagent protein. For better results, it is recommended that immunological assays be performed as soon as possible.

In addition, the inefficacy of Wondfo® IC test in detected SARS-CoV-2 IgG or IgM could be associated to the timing of sampling. The COVID-19 disease has different infections stages, which the immunological response vary with the disease pathology [13M14]. Studies have been demonstrated that the timing of sampling reflects in the sensitivity and specificity of serological tests. Imai, et al. [15] show that the sensitivity of IC assay was low during the early phase in symptomatic and asymptomatic patients. Cassaniti, et al. [16] reported that the VivaDiag® COVID-19 IgM/IgG Rapid Test present a weak positive serology in acute patients. In this study the rapid test detected just 18.4% of positive samples of patients confirmed to be positive for COVID-19 by real time RT-PCR. This contradiction could be related to the timing of sampling because it reflects the develop of the disease. In this context, Pan, et al. [8] suggest that that the sensitivity of IC assay fluctuate with the disease progression, the authors shows that the positive rates of SARS-CoV-2 IgM or IgG in the early stage are relatively low, and gradually increase during the disease evolution. So, the stage of the infection, which was reflected in the timing of sampling, could infer in the results of an immunochromatographic assay.

This study has some limitations. The number of serological samples and the rapid tests performed. The use of only one serological methodology as a reference for the presence of antibodies. In addition, the positive and negative samples distribution could be a limitation of the study.

Thus, the Wondfo® IC test failed in providing a quick, valid and reliable results and appears not to be a good alternative for clinical use in detecting pandemic coronavirus. However, if the limitations of the rapid test are known, some correction factors can be used in order to adjust the epidemiological data. Further studies may be necessary to determine the usefulness of this kit in different settings and communities. To further understand the reasons behind the lower sensitivity of the immunochromatographic assay, future research should be designed to investigate the possibility of interferences of sample timing and collection in the kit result.

This work was supported by grants from FAPITEC-SE, CNPq and CAPES, all agencies from Brazil. We thank the health department of the state of Sergipe for donating the tests, to the SOLIM, CLIMED, CIRURGIA laboratories for the interlaboratory evaluation. This work was supported by grants from the Brazilian agencies, CAPES, CNPq and FAPITEC/SE. The authors dedicate this article and thank all the doctors and frontline health workers and other staff for their so much dedication to facing COVID-19.

Contribution

LPB, LQJ and AASA contributed to conception and design of the study. LPB, MGBO, DR and KAS contributed to the acquisition of data and analysis of data. LPB, LQJ and AASA contributed to analysis and interpretation of the data. LH, LPB and LQJ drafting the article or revising it critically for important intellectual content. LPB, MGBO, DR, LH, LQJ and AASA approved the version to be submitted.

  1. Huang C, Wang Y, Li X, Ren L, Zhao J, et al. Clinical features of patients infected with 2019 novel coronavirus in Wuhan, China. Lancet. 2020; 395: 497–506.
  2. WHO. WHO Coronavirus disease 2019 (COVID-19): situation report, accessed may18, 2020. https://www.who.int/emergencies/diseases/novel-coronavirus  
  3. Jin Y, Yang H, Ji W, Wu W, Chen S, et al. Virology, Epidemiology, Pathogenesis, and Control of COVID-19. Viruses. 2020; 27: 12: 372. PubMed: https://pubmed.ncbi.nlm.nih.gov/32230900/  
  4. Ahn DG, Shin HJ, Kim MH, Lee S, Kim HS, et al. Current Status of Epidemiology, Diagnosis, Therapeutics, and Vaccines for Novel Coronavirus Disease 2019 (COVID-19). J Microbiol Biotechnol. 2020; 30: 313–324. PubMed: https://pubmed.ncbi.nlm.nih.gov/32238757/
  5. Vashist SK. In Vitro Diagnostic Assays for COVID-19: Recent Advances and Emerging Trends. Diagnostics (Basel). 2020; 10: 202. PubMed: https://pubmed.ncbi.nlm.nih.gov/32260471/
  6. Loeffelholz MJ, Tang Y-W. Laboratory diagnosis of emerging human coronavirus infections - the state of the art. Emerg Microbes Infect. 2020; 9: 747–756. PubMed: https://pubmed.ncbi.nlm.nih.gov/32196430/
  7. Tang YW, Schmitz JE, Persing DH, Stratton CW. The Laboratory Diagnosis of COVID-19 Infection: Current Issues and Challenges. J Clin Microbiol. 2020; 58: 512-520. PubMed: https://pubmed.ncbi.nlm.nih.gov/32245835/
  8. Pan Y, Li X, Yang G, Fan J, Tang Y, et al. Serological immunochromatographic approach in diagnosis with SARS-CoV-2 infected COVID-19 patients. J Infect. 2020; 81: 28-32. PubMed: https://pubmed.ncbi.nlm.nih.gov/32283141/
  9. Montesinos I, Gruson D, Kabamba B, Dahma H, Van den Wijngaert S, et al. Evaluation of two automated and three rapid lateral flow immunoassays for the detection of anti-SARS-CoV-2 antibodies. J Clin Virol. 2020; 128: 104413. PubMed: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7198434/
  10. Venter M, Richter K. Towards effective diagnostic assays for COVID-19: a review. J Clin Pathol. 2020; 73: 370-377. PubMed: https://pubmed.ncbi.nlm.nih.gov/32404473/  
  11. Demey B, Daher N, François C, Lanoix JP, Duverlie G, et al. Dynamic profile for the detection of anti-SARS-CoV-2 antibodies using four immunochromatographic assays. J Infect. 2020; PubMed: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7204722/
  12. Vásárhelyi B, Kristóf K, Ostorházi E, Szabó D, Prohászka Z, et al. The diagnostic value of rapid anti IgM and IgG detecting tests in the identification of patients with SARS CoV-2 virus infection. Orv Hetil. 2020; 161: 807–812. PubMed: https://pubmed.ncbi.nlm.nih.gov/32364358/
  13. Chen G, Wu D, Guo W, Cao Y, Huang D, et al. Clinical and immunological features of severe and moderate coronavirus disease 2019. J Clin Invest. 2020; 130: 2620–2629. PubMed: https://pubmed.ncbi.nlm.nih.gov/32217835/
  14. Vardhana SA, Wolchok JD. The many faces of the anti-COVID immune response. J Exp Med. 2020; 217: 20200678. PubMed: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7191310/
  15. Imai K, Tabata S, Ikeda M, Noguchi S, Kitagawa Y, et al. Clinical evaluation of an immunochromatographic IgM/IgG antibody assay and chest computed tomography for the diagnosis of COVID-19. J Clin Virol. 2020; 128: 104393. PubMed: https://pubmed.ncbi.nlm.nih.gov/32387968/  
  16. Cassaniti I, Novazzi F, Giardina F, Salinaro F, Sachs M, et al. Members of the San Matteo Pavia COVID-19 Task Force. Performance of VivaDiag COVID-19 IgM/IgG Rapid Test is inadequate for diagnosis of COVID-19 in acute patients referring to emergency room department. J Med Virol. 2020; 10. PubMed: https://pubmed.ncbi.nlm.nih.gov/32227490/